Accuracy of isoflurane, halothane, and sevoflurane vaporizers during high oxygen flow and at maximum vaporizer dial
Tamas D Ambrisko1, Alan M Klide
1Department of Clinical Studies, School of Veterinary Medicine, University of Pennsylvania, Philadelphia, PA 19104, USA. tamas.ambrisko@vetmeduni.ac.at
Objective:
To assess the accuracy of isoflurane, halothane, and sevoflurane vaporizers during high oxygen flow and at maximum dial settings at room temperature and to test sevoflurane vaporizers similarly during heating and at low-fill states.
Sample:
5 isoflurane, 5 halothane, and 5 sevoflurane vaporizers.
Procedures:
Vaporizers were tested at an oxygen flow of 10 L/min and maximum dial settings for 15 minutes under various conditions. All 3 vaporizer types were filled and tested at room temperature (21° to 23°C). Filled sevoflurane vaporizers were wrapped with circulating hot water (42°C) blankets for 2 hours and tested similarly, and near-empty sevoflurane vaporizers were tested similarly at room temperature. During each 15-minute test period, anesthetic agent concentration was measured at the common gas outlet with a portable refractometer and temperature of the vaporizer wall was measured with a thermistor.
Results:
For each vaporizer type, anesthetic agent concentrations and vaporizer wall temperatures decreased during the 15-minute test period. Accuracy of isoflurane and halothane vaporizers remained within the recommended 20% (plus or minus) deviation from dial settings. Heated and room-temperature sevoflurane vaporizers were accurate to within 23% and 11.7% (plus or minus) of dial settings, respectively. Sevoflurane vaporizers at low-fill states performed similarly to vaporizers at full-fill states.
Conclusions And Clinical Relevance:
Under these study conditions, the isoflurane and halothane vaporizer models tested were accurate but the sevoflurane vaporizers were not. Sevoflurane vaporizer accuracy was not affected by fill state but may be improved with vaporizer heating; measurements of inspired anesthetic agent concentrations should be obtained during the use of heated vaporizers.
More Related Videos
06:27The Serial Anesthesia Array for the High-Throughput Investigation of Volatile Agents Using Drosophila melanogaster
Published on: February 24, 2023
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Inhalational Anesthetics: Overview
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Vapor Pressure of Fluid
When a liquid is placed in a closed container with a small air space, and the space is evacuated, vapor molecules will...
Vapor Pressure
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.