Analytical Techniques for Assaying Nitric Oxide Bioactivity
Published on: June 18, 2012
Can we use portable nitric oxide analyzer in young children?
Satu Kalliola1, Pekka Malmberg, Tuija Rito
1Department of Allergy, Helsinki University Central Hospital, Helsinki, Finland. satu.kalliola@hus.fi
The portable NIOX MINO device shows promise for screening airway inflammation in children aged 4+, but has lower success rates and accuracy at low fractional exhaled nitric oxide (FENO) levels compared to the stationary NIOX device.
Area of Science:
- Pediatric Pulmonology
- Respiratory Medicine
- Biomedical Engineering
Background:
- Asthma management can be enhanced by measuring fractional exhaled nitric oxide (FENO).
- Portable FENO analyzers offer potential for primary care settings.
- Previous studies validated devices in adults and older children, but not young children.
Purpose of the Study:
- To compare the accuracy and usability of a portable electrochemical device (NIOX MINO) with a stationary chemiluminescence device (NIOX) for FENO measurement in young children.
- To assess the feasibility of using NIOX MINO in primary care for pediatric asthma assessment.
Main Methods:
- Paired FENO measurements were conducted using stationary NIOX and portable NIOX MINO devices.
- Participants included children with asthmatic symptoms and age-matched healthy controls.
- Data analysis focused on success rates, reproducibility, correlation, and agreement between devices.
Main Results:
- The NIOX MINO device had a lower success rate (73%) compared to NIOX (93%) in children aged 3.9-8.5 years.
- Both devices demonstrated similar reproducibility and a strong correlation (r=0.97).
- Despite correlation, Bland-Altman analysis revealed wide limits of agreement, and both devices distinguished higher FENO in asthmatic children.
Conclusions:
- The NIOX MINO can serve as a screening tool for airway inflammation in children aged 4 and above.
- Its clinical utility is constrained by a lower measurement success rate and limited accuracy at low FENO levels.
- Further validation is needed to optimize the use of portable FENO devices in young pediatric populations.