Comparison of cyclodeviation and duction measurement in Graves' orbitopathy patients using different devices
Hinke Marijke Jellema1, Annette Baader, Susanne Pitz
1Department of Ophthalmology, Orthoptic Unit, Academic Medical Center, Amsterdam, The Netherlands. h.m.jellema@amc.uva.nl
Purpose:
To compare measurement outcomes of different devices measuring cyclodeviation and ductions in Graves' orbitopathy (GO) patients.
Methods And Materials:
Cyclodeviation in GO patients was measured using the Harms tangent screen (HTS), the cycloforometer of Franceschetti, and the synoptometer. Ductions were measured using the modified perimeter, the Goldmann perimeter and the Maddox tangent scale (MTS).
Results:
In 13 patients, cyclodeviation in primary position, up-, and downgaze was measured with the above-mentioned devices. The mean differences ranged from 0.3º to 3.1º and were smallest between the HTS and the cycloforometer (89% of all measurements within 2º difference). Measurement of abduction, adduction, elevation, and depression using the modified perimeter, Goldmann perimeter, and MTS were obtained in another 13 patients. The mean differences ranged from 1.2º to 12.9º and were smallest between the modified perimeter and the Goldmann perimeter (92% of all measurements ≤ 8º).
Conclusions:
The HTS and cycloforometer produce interchangeable measurement outcomes. The modified perimeter and the Goldmann perimeter are interchangeable as well. However, the synoptometer and the MTS are not suitable for comparative analysis.
Related Concept Videos
Graves' Disease I: Introduction
Graves Disease II: Pathophysiology
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...