Preventing overdiagnosis of implantable cardioverter-defibrillator lead fractures using device diagnostics
Charles D Swerdlow1, Haresh Sachanandani, Bruce D Gunderson
1The Cedars-Sinai Heart Institute, Cedars-Sinai Medical Center, Los Angeles, CA, USA. swerdlow@ucla.edu
Objectives:
This study sought to use implantable cardioverter-defibrillator (ICD) diagnostics to discriminate ICD lead fractures from normally functioning leads with high impedance and from connection problems between the lead and header.
Background:
ICD diagnostics facilitate identification of fractures, but there are no accepted criteria for discriminating fractures from other causes of high impedance and/or nonphysiological "noise" oversensing.
Methods:
We analyzed a development set of 91 leads to construct a stepwise algorithm based on ICD diagnostics. It included 40 fractures, 30 connection problems, and 21 functioning leads that triggered high-impedance alerts. Then we applied this algorithm to an independent test set of 100 leads: 70 fractures and 30 intact leads with connection problems that were misdiagnosed clinically as fractures. In the algorithm, either extremely high maximum impedance or noise oversensing with a normal impedance trend indicated a fracture. A short interval from surgery to impedance rise or prolonged stable impedance after an abrupt rise indicated a connection problem. A gradual impedance increase or stable, high impedance indicated a functioning lead.
Results:
In the test set, the algorithm correctly classified 100% of fractures (95% confidence interval [CI]: 95% to 100%) and 87% of connection problems that were misdiagnosed as fractures (95% CI: 70% to 95%).
Conclusions:
An algorithm using only ICD diagnostics identifies leads with oversensing or high impedance as fractures or connection problems with a high degree of accuracy.