Related Experiment Video
Updated: Jun 1, 2026

A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
Biomechanical comparison of three fixation techniques used for four-corner arthrodesis
J Kraisarin1, D G Dennison, L J Berglund
1Biomechanics Laboratory, Division of Orthopedic Research, Mayo Clinic College of Medicine, Rochester, USA.
Abstract:
Clinical results following four-corner arthrodesis vary and suggest that nonunion may be related to certain fixation techniques. The purpose of our study was to examine the displacement between the lunate and capitate following a simulated four-corner arthrodesis with the hypothesis that three types of fixation (Kirschner wires, dorsal circular plate, and a locked dorsal circular plate) would allow different amounts of displacement during simulated wrist flexion and extension. Cadaver wrists with simulated four-corner arthrodeses were loaded cyclically either to implant failure or until the lunocapitate displacement exceeded 1 mm. The locked dorsal circular plate group was significantly more stable than the dorsal circular plate and K-wire groups (p = 0.018 and p = 0.006). While these locked dorsal circular plates appear to be very stable our results are limited only to the biomechanical behavior of these fixation techniques within a cadaver model.