High-resolution mass spectrometry with a multiple pass quadrupole mass analyzer
1Ames Laboratory [Formula: see text] U.S. Department of Energy, Department of Chemistry, Iowa State University, Ames Iowa 50011.
Analytical Chemistry
|June 8, 2011
Abstract:
The peak shape narrows and the resolution improves if the ions are simply reflected back and forth through a conventional quadrupole mass analyzer. CO(+) and N(2)(+) at m/z = 28 are separated to 50% valley with half of the original signal remaining. These two ions can be resolved to baseline (m/Δm) = 5000 with 1% of the original signal remaining.
Related Concept Videos
Mass Analyzers: Common Types
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
Mass Analyzers: Overview
The mass analyzer is a crucial component of the mass spectrometer. In the ionization chamber, the vaporized sample is bombarded with a high-energy electron beam to generate a radical cation and further fragment into neutral molecules, radicals, and cations. A series of negatively charged accelerator plates accelerate the cations into the mass analyzer. The mass analyzer separates ions according to their mass-to-charge (m/z) ratios and then directs them to the detector. The common types of mass...
High-Resolution Mass Spectrometry (HRMS)
The resolution of a mass spectrometer depends on the efficiency of separating ions with different ion masses. The mass of an atom is approximated to the sum of the masses of protons and neutrons inside, considering the masses of protons and neutrons as equal. However, the masses of the proton (1.6726 × 10−24 g) and neutron (1.6749 × 10−24 g) are not truly equal. There is a minor error in the expression of atomic masses relative to the simplest atom of hydrogen. For example, the mass of helium...
Mass Spectrometry: Complex Analysis
Mass spectrometry is an important technique for the identification of pure compounds. However, it has some limitations for the analysis of complex mixtures, often due to excessive fragmentation making the spectrum too complicated to decipher. Mass spectrometry can be combined with suitable separation methods in sequence, forming hyphenated methods, which are useful in the analysis of complex mixtures.
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrum: Interpretation
An unknown compound can be established by identifying the molecular ion peak in the mass spectrum. The molecular ion peak is often weak or absent due to the predominance of fragmentation in high-energy electron beams. In such cases, a soft-energy electron beam can be used to scan the spectrum to enhance the intensity of the molecular ion peak. Additionally, chemical ionization, field ionization, and desorption ionization spectra are used to obtain a relatively intense molecular ion peak.To...
Tandem Mass Spectrometry
Tandem mass spectrometry is a technique that uses multiple mass analyzers in series to obtain a higher selectivity and reduce chemical noise during analyte detection. Instruments with multiple analyzers separated by an interaction cell enable secondary fragmentation and selected study of the fragment ions.Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called...