Related Experiment Video
Updated: Jun 1, 2026

Non-equilibrium Microwave Plasma for Efficient High Temperature Chemistry
Published on: August 1, 2017
A graphite furnace electrothermal vaporization system for inductively coupled plasma atomic emission spectrometry
1Sektion Analytik und Höchstreinigung, Universität Ulm, D-89069 Ulm, Germany.
Abstract:
An improved graphite furnace electrothermal vaporization device equipped with an autosampler for inductively coupled plasma atomic emission spectrometry is presented. The transport losses of eight selected analytes in the individual segments of the device were determined by means of the radiotracer technique by applying amounts traced comparable to those to be determined in real samples. The results obtained from the radiotracer study were the basis for further improvement of the interface design, leading to considerable increase of the total transport efficiency, which finally was found to be between 26 (for Cr) and 57% (for Ga). The whole system consists of a graphite furnace vaporizer, a power supply, a gas flow box, and an autosampler with incorporated microbalance. The temperature program, gas flows, and autosampler functions are controlled by a data station which also provides the data acquisition and processing of the transient signals. The performance parameters of the developed system were evaluated using aqueous standard solutions. Absolute limits of detection for most analytes were between 0.1 and 1 ng, and for As, K, Ni and Pb, they were between 2 and 3.2 ng.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview