Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Muscle Stimulation Frequency01:22

Muscle Stimulation Frequency

The contraction strength of muscles is regulated by motor neurons, which modulate the frequency of action potentials dispatched to the motor units based on the body's requirements. This process of varying the muscle stimulation frequency allows muscles to contract with a force that is precisely tailored to the needs of the moment, whether lifting a feather or a heavy box.
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...
Interference: Path Lengths01:10

Interference: Path Lengths

Consider two sources of sound, that may or may not be in phase, emitting waves at a single frequency, and consider the frequencies to be the same.
Two special sources may be considered when they are in phase. This can be easily achieved by feeding the two sources from the same source. An example would be synchronizing the two speakers by feeding them with the same source, such as the sound waves produced by a tuning fork. This setup ensures that the two sources have the same frequency and are...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Grading quality score of contrast electrical impedance tomography for lung ventilation-perfusion: development and preliminary evaluation of the impedance-time curve.

Respiratory medicine·2026
Same author

Thoracic electrical impedance tomography using a two-plane electrode configuration in horses.

Equine veterinary journal·2026
Same author

A robust temporal metric of ventilation inhomogeneity in electrical impedance tomography.

Physiological measurement·2026
Same author

Lung perfusion estimation by saline-contrast EIT without breath hold: a randomized cross-over trial.

Critical care (London, England)·2026
Same author

Comparison of early and late time-controlled adaptive ventilation on pulmonary gas exchange in anesthetized horses.

Veterinary anaesthesia and analgesia·2025
Same author

Bedside assessment of lung ventilation, recruitability and optimal PEEP by 3D-EIT and 2D-EIT in ARDS patients: a prospective study.

Intensive care medicine·2025
Same journal

Continuous tracking of aortic aneurysm diameter with peripheral pulse waves: a computational framework combining sequential Markov chain Monte Carlo with Kalman filtering.

Physiological measurement·2026
Same journal

The 2026 global roadmap for textile-integrated wearable technologies in health.

Physiological measurement·2026
Same journal

Augmenting single-lead ECG interpretation through QRS waveform decomposition and rotation.

Physiological measurement·2026
Same journal

Dynamic Beat-to-Beat Blood Pressure Estimation using a Multi-modal Wearable Deep Learning Approach.

Physiological measurement·2026
Same journal

Dual warm-start fusion versus attention-based fusion in low-label ECG-PCG classification: a controlled ablation study.

Physiological measurement·2026
Same journal

Inter-patient multi-label ECG classification via low-rank adaptation fine-tuned large language models with dynamic graph convolutional network.

Physiological measurement·2026
See all related articles

Related Experiment Video

Updated: Jun 1, 2026

Automated Multimodal Stimulation and Simultaneous Neuronal Recording from Multiple Small Organisms
08:28

Automated Multimodal Stimulation and Simultaneous Neuronal Recording from Multiple Small Organisms

Published on: March 3, 2023

Adjacent stimulation and measurement patterns considered harmful.

Andy Adler1, Pascal Olivier Gaggero, Yasheng Maimaitijiang

  • 1Systems and Computer Engineering, Carleton University, Ottawa, Canada. adler@sce.carleton.ca

Physiological Measurement
|June 8, 2011
PubMed
Summary
This summary is machine-generated.

Electrical impedance tomography (EIT) can better detect internal conductivity changes using improved electrode patterns. Adjacent patterns perform poorly; trigonometric or separated pair patterns significantly enhance imaging performance.

More Related Videos

Concurrent Electroencephalography Recording During Transcranial Alternating Current Stimulation (tACS)
06:51

Concurrent Electroencephalography Recording During Transcranial Alternating Current Stimulation (tACS)

Published on: January 22, 2016

Related Experiment Videos

Last Updated: Jun 1, 2026

Automated Multimodal Stimulation and Simultaneous Neuronal Recording from Multiple Small Organisms
08:28

Automated Multimodal Stimulation and Simultaneous Neuronal Recording from Multiple Small Organisms

Published on: March 3, 2023

Concurrent Electroencephalography Recording During Transcranial Alternating Current Stimulation (tACS)
06:51

Concurrent Electroencephalography Recording During Transcranial Alternating Current Stimulation (tACS)

Published on: January 22, 2016

Area of Science:

  • Biomedical Engineering
  • Medical Imaging Physics

Background:

  • Electrical impedance tomography (EIT) is a non-invasive imaging technique.
  • EIT reconstructs internal conductivity distributions from surface measurements.
  • Optimizing stimulation and measurement patterns is crucial for EIT performance.

Purpose of the Study:

  • To characterize the distinguishability of internal conductivity changes using EIT.
  • To analyze the impact of stimulation and measurement patterns on EIT performance.
  • To propose an improved EIT imaging protocol.

Main Methods:

  • Development of a distinguishability measure (z) based on signal-to-noise ratio and probability of detection.
  • Analysis of various stimulation and measurement patterns under medical electrical safety constraints.
  • Simulations for 3D tank and chest shapes and experimental validation in a saline tank with 16 electrodes.

Main Results:

  • Adjacent stimulation and measurement patterns exhibit the poorest performance, being 6.9 times less effective.
  • Trigonometric patterns and pair drive/measurement patterns separated by over 90° show significantly improved results.
  • A new measure 'z' quantifies the impact of pattern selection on conductivity change detection.

Conclusions:

  • Adjacent EIT patterns are suboptimal and should be abandoned due to poor performance.
  • Pair drive and measurement patterns separated by approximately 180° are recommended for improved EIT imaging.
  • Modifications to existing EIT systems, including electrode placement adjustments, can enhance performance.