Related Experiment Video
Updated: Jun 1, 2026

Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
A heparin-coated dialysis filter (AN69 ST) does not reduce clotting during hemodialysis when compared to a
Solbjørg Sagedal1, Bartlomiej J Witczak, Kåre Osnes
1Department of Nephrology, Oslo University Hospital Ullevål, Oslo, Norway.
Background:
We investigated whether the heparin-coated AN69 ST hemodialysis (HD) filter induced less hypercoagulability during HD than a conventional polysulfone filter (F×8).
Methods:
In a crossover design, 11 patients were treated alternately with AN69 ST and F×8 filters (45 sessions). All filters were primed with unfractionated heparin (UFH) and unadsorbed UFH was removed by saline flushing. Half the conventional dalteparin dose was given as a bolus dose at the start of HD. Clotting was evaluated hourly in the venous air trap. Prothrombin fragments 1 and 2 (PF1 + 2), antithrombin (AT), β-TG and anti-FXa activity were repeatedly measured.
Results:
One patient treated with enalapril had two repeated adverse reactions to the AN69 ST filter and was excluded from the study. Use of the AN69 ST filter did not decrease the mean clot score or PF1 + 2, but decreased β-TG compared to the F×8 filter.
Conclusion:
The heparin-coated AN69 ST filter did not induce less coagulation when compared to the F×8 filter.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Hemodialysis I: Introduction
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Anticoagulant Drugs: Low-Molecular-Weight Heparins