Related Experiment Video
Updated: Jun 1, 2026

The Flexible Rhino-Laryngoscope for Awake Nasotracheal Intubation
Published on: August 2, 2024
Indirect versus direct laryngoscopy for routine nasotracheal intubation
Wolfgang Puchner1, Lukas Drabauer, Klemens Kern
1Department of Anesthesia and Intensive Care, General Hospital Linz, A-4020, Linz, Austria. wolfgang.puchner@akh.linz.at
Study Objective:
To compare the effectiveness of the indirect laryngoscopes, Airtraq (A) and GlideScope (G), with the Macintosh (M) laryngoscope in routine nasotracheal intubation.
Design:
Randomized, single-blinded study.
Setting:
University-affiliated, tertiary-care hospital.
Patients:
62 adult, ASA physical status 1 and 2 patients with normal airways requiring nasotracheal intubation for dental or maxillofacial surgery.
Intervention:
Patients in Groups A and G underwent nasal intubation with the Airtraq and GlideScope, respectively, while laryngoscopy in Group M was performed with the Macintosh blade.
Measurements:
Performance of the intubating tools was judged by the ease [Intubation Difficulty Scale (IDS) and numeric rating scale (NRS)] and time to intubation (laryngoscopy and endotracheal tube advancement). In addition, hemodynamic parameters, severity of postoperative sore throat, and posture of the intubator were recorded.
Main Results:
IDS score was significantly lower with the Airtraq and GlideScope than with the Macintosh laryngoscope (mean ± SD: A 0.1 ± 0.3, G 0.3 ± 0.6, M 0.8 ± 1.0; P = 0.013). NRS reported by the intubators showed a similar preference for indirect over direct laryngoscopy (A 0.9 ± 0.7, G 1.1 ± 0.6, M 1.9 ± 1.1; P = 0.001). Duration of laryngoscopy and endotracheal tube insertion was similar in all groups. No significant intergroup differences in hemodynamic parameters were recorded. Postoperative sore throat was significantly reduced using the GlideScope compared with the other devices (P = 0.048).
Conclusion:
The Airtraq and GlideScope facilitated nasotracheal intubation more so than the Macintosh laryngoscope in adults with apparently normal airways.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: