High resolution at low beam energy in the SEM: resolution measurement of a monochromated SEM
1Sandia National Laboratories, Materials Characterization Department, Albuquerque, New Mexico, USA. jrmicha@sandia.gov
Abstract:
The resolution of secondary electron low beam energy imaging of a scanning electron microscope equipped with a monochromator is quantitatively measured using the contrast transfer function (CTF) method. High-resolution images, with sub-nm resolutions, were produced using low beam energies. The use of a monochromator is shown to quantitatively improve the resolution of the SEM at low beam energies by limiting the chromatic aberration contribution to the electron probe size as demonstrated with calculations and images of suitable samples. Secondary electron image resolution at low beam energies is ultimately limited by noise in the images as shown by the CTFs.