Real-time assessment of myocardial contractility using shear wave imaging
Mathieu Pernot1, Mathieu Couade, Philippe Mateo
1Institut Langevin, École Supérieure de Physique et de Chimie Industrielles, ParisTech, 10 rue Vauquelin, Paris, France. mathieu.pernot@espci.fr
Objectives:
The goal of this study was to assess whether myocardial stiffness could be measured by shear wave imaging (SWI) and whether myocardial stiffness accurately quantified myocardial function.
Background:
SWI is a novel ultrasound-based technique for quantitative, local, and noninvasive mapping of soft tissue elastic properties.
Methods:
SWI was performed in Langendorff perfused isolated rat hearts (n = 6). Shear wave was generated and imaged in the left ventricular myocardium using a conventional ultrasonic probe connected to an ultrafast scanner (12,000 frames/s). The local myocardial stiffness was derived from shear wave velocity every 7.5 ms during 1 single cardiac cycle.
Results:
The average myocardial stiffness was 8.6 ± 0.7 kPa in systole and 1.7 ± 0.8 kPa in diastole. Myocardial stiffness was compared with isovolumic systolic pressure at rest and during administration of isoproterenol (10(-9), 10(-8), and 10(-7) mol/l, 5 min each). Systolic myocardial stiffness increased strongly up to 23.4 ± 3.4 kPa. Myocardial stiffness correlated strongly with isovolumic systolic pressure (r(2) = [0.94; 0.98], p < 0.0001).
Conclusions:
Myocardial stiffness can be measured in real time over the cardiac cycle using SWI, which allows quantification of stiffness variation between systole and diastole. Systolic myocardial stiffness provides a noninvasive index of myocardial contractility.