Related Experiment Videos
Verification of an electromagnetic placement device compared with abdominal radiograph to predict accuracy of feeding
Jan Powers1, Michael Luebbehusen, Tracy Spitzer
1St Vincent Hospital, Indianapolis, Indiana, USA. jmpowers@stvincent.org
Background:
Use of an electromagnetic placement device (EMPD) facilitates placement of feeding tubes at the bedside. Standard practice for verification of feeding tube placement is via radiographic confirmation. The purpose of this research study was to assess the accuracy of placement of small-bore feeding tubes (SBFTs) as determined by EMPD interpretation compared with that of abdominal radiograph verification by a radiologist.
Methods:
This multicenter prospective study enrolled patients requiring bedside feeding tube placement. SBFTs were placed by an experienced investigator using the EMPD. Two abdominal radiographs were then obtained: one after initial SBFT placement and an additional radiograph after injection of contrast. Documentation of location based on clinician interpretation using the EMPD was then compared with radiologist interpretation.
Results:
The final sample size was 194 patients, including 18 pediatric patients. Patient age ranged from 12 days to 102 years. Median time for tube placement was 12 minutes. Of the 194 patients, only 1 patient had data showing discrepancies between the original EMPD verification and the final abdominal radiograph interpretation, providing a 99.5% agreement. No patient experienced complications during SBFT placement, and 15 patients had inadvertent airway placement that was avoided with the use of the EMPD.
Conclusions:
There was a high percentage of agreement between EMPD and radiologic interpretation after contrast injection. The EMPD aided in avoiding inadvertent airway placement, with no patient complications. This device can be used safely at the bedside to facilitate placement of feeding tubes, leading to the delivery of early enteral nutrition.
Related Concept Videos
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...