Lumen claims of the STERRAD 100NX sterilizer: testing performance limits when processing equipment containing long,
Magda Diab-Elschahawi1, Alexander Blacky, Nicole Bachhofner
1Clinical Institute of Hospital Hygiene, Vienna General Hospital, Medical University of Vienna, Austria. magda.diab-elschahawi@meduniwien.ac.at
Background:
According to manufacturers information, the STERRAD 100NX sterilizer-a low temperature H(2)O(2) gas plasma sterilizer-can adequately process single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a maximum length of 500 mm using standard cycle sterilizing conditions. The aim of this study was to qualify the performance of this H(2)O(2) gas plasma sterilizer under different experimental settings representing worst case conditions.
Methods:
Inoculated carriers were placed at the midpoint position of specified lumens and then submitted to flex scope sterilizing conditions. To simulate insufficient cleaning or crystalline residues, we added organic and inorganic challenges to our inoculated carriers.
Results:
For experiments done with unchallenged carriers, quantitative analysis reached a log(10) reduction rate of ≥5.71, whereas qualitative results showed no growth in 24 out of 30 biologic indicators tested using flex scope half cycle conditions. Any additional kind of challenge significantly impaired the sterilization outcome.
Conclusion:
The findings of our current study emphasize the importance of a thorough validated cleaning of medical devices as well as timing for cleaning and decontamination before being exposed to the H(2)O(2) sterilization process and, furthermore, the need for strict adherence to manufacturer's recommendations.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Scale-Up Processes
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...