Performance of the On-Q pain infusion device during changes in environmental temperature
Tyler Burnett1, Tarun Bhalla, Amod Sawardekar
1Department of Anesthesiology, Nationwide Children's Hospital and Ohio State University, Columbus, OH 43205, USA.
Background:
The On-Q infusion device is an elastomeric device with a flow regulator that controls the flow of a local anesthetic agent through a peripheral catheter. As variations in external temperature may affect the diameter of the tubing or viscosity of the fluid, it is feasible that alterations in flow may be caused by such temperature variations. This study evaluates the performance of this device during variations in environmental temperature.
Methods:
The disposable 400 ml On-Q pain ball infusion devices were filled and connected to a single, end hole infusion catheter and set to infuse at 14 ml·h(-1) . Eighteen devices were used in the study (six of each at three different temperatures). The temperatures included hot (54°C), room temperature (21°C), and cold (6°C). The devices were allowed to flow for 24 h. The fluid delivered during each 12-h period was measured using a graduated column.
Results:
There were significant differences in the output from the devices at the hot (54°C) temperature and the cold (6°C) temperature when compared to room temperature (21°C). When compared to room temperature, the output decreased to 67% and 54% of the control group (room temperature) during hours 0-12 and 12-24, respectively, in a cold environment (6°C). An increased external temperature resulted in a greater output from the devices. When compared to the room temperature devices, the output was 49% higher during the first 12 h and 40% higher during the second 12 h at an external temperature of 54°C.
Conclusions:
This preliminary investigation demonstrates what may be clinically significant changes in output from the On-Q pain device based on the external temperature. These alterations in flow could result in inadequate analgesia or even potentially toxicity if these devices are used in smaller patients especially the pediatric population.
Related Concept Videos
Decreased Body Temperature
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Factors Affecting Body Temperature
Factors may include:
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...