Related Experiment Videos
A comparison of four immunometric assays for myeloperoxidase using luminescent and colorimetric signal detection
1Institute of Microbiology, Czechoslovak Academy of Science, Prague, CSFR.
Abstract:
This communication presents four different assay systems for the determination of myeloperoxidase in body fluids. One is based on conventional chemiluminescence, two on luminescence-amplified enzyme measurement using either spiroadamantane-1,2-dioxetanes with alkaline phosphatase or luminol/peroxidase/4-iodophenol coupled with a peroxidase label. The assays covered the range 0-600 micrograms/l peroxidase. Established reference range for EDTA plasma from healthy volunteers gave an upper limit of 250 micrograms/l (95% confidence limits). Intra-assay coefficients of variation were less than 5% for all assays, as seen in compound precision profiles. Inter-assay variation was less than 10% throughout the whole concentration range. Kinetic curves for the light production were performed over a 2 hour period for the enhanced luminescence assays. Dynamic ranges of these assays were compared with the conventional colorimetric assay using 4-nitrophenyl phosphate--alkaline phosphatase. The assay was standardized using a commercially available myeloperoxidase preparation with defined enzymatic activity. The protein content was estimated and the laboratory standard compared and calibrated in mass units.