Changes in reference vessel diameter in ST-segment elevation myocardial infarction after primary percutaneous
Ecaterina Cristea1, Gregg W Stone, Roxana Mehran
1Columbia University Medical Center and the Cardiovascular Research Foundation, New York, NY, USA.
American Heart Journal
|July 12, 2011
Summary
Stent undersizing is not the cause of stent thrombosis in ST-segment elevation myocardial infarction (STEMI). Reference vessel diameter remains stable after percutaneous coronary intervention (PCI) regardless of stent type, indicating appropriate sizing.
Area of Science:
- Cardiology
- Interventional Cardiology
- Biomedical Engineering
Background:
- Stent undersizing during primary percutaneous coronary intervention (PCI) for ST-segment elevation myocardial infarction (STEMI) may increase stent thrombosis.
- Reference vessel diameter (RVD) is a critical factor in stent selection and outcomes.
Purpose of the Study:
- To compare RVD of the infarct-related artery before, immediately after, and at late follow-up after PCI for STEMI.
- To investigate differences in RVD behavior between paclitaxel-eluting stents (PES) and bare-metal stents (BMS) at follow-up.
Main Methods:
- Analysis of quantitative angiographic data from 2,974 patients (3,589 lesions) in the HORIZONS-AMI trial.
- Comparison of RVD at baseline, post-PCI, and 13-month follow-up for lesions treated with PES (2,233) and BMS (741).
Main Results:
- Median RVD was stable: 2.87 mm (baseline), 2.92 mm (post-PCI), and 2.88 mm (follow-up).
- No significant differences in RVD were observed between PES and BMS groups at any time point.
- RVD at baseline and post-PCI was similar in patients with and without stent thrombosis.
Conclusions:
- Reference vessel diameter does not significantly change from baseline to follow-up after PCI for STEMI.
- Stent sizing appears appropriate for vessel size, and stent type does not influence RVD changes.
- Stent undersizing is unlikely to be the primary driver of high stent thrombosis rates in STEMI.