Related Experiment Video
Updated: May 31, 2026

Synthesis and Operation of Fluorescent-core Microcavities for Refractometric Sensing
Published on: March 13, 2013
Absolute refractometry of dry gas to ±3 parts in 10⁹
1National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, Maryland 20899, USA.
Abstract:
We present a method of measuring the refractive index of dry gases absolutely at 632.8 nm wavelength using a Fabry-Perot cavity with an expanded uncertainty of <3×10⁻⁹ (coverage factor k=2). The main contribution to this uncertainty is how well vacuum-to-atmosphere compression effects (physical length variation) in the cavities can be corrected. This paper describes the technique and reports reference values for the refractive indices of nitrogen and argon gases at 100 kPa and 20 °C with an expanded uncertainty of <9×10⁻⁹ (coverage factor k=2), with the additional and larger part of this uncertainty coming from the pressure and temperature measurement.
Related Concept Videos
Deviation from Ideal Behaviour
Gas Solubility
Real Gases: Effects of Intermolecular Forces and Molecular Volume Deriving Van der Waals Equation
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Nonideal Two-Component Liquid Solutions