Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: October 31, 2010
The functioning of oxygen concentrators in resource-limited settings: a situation assessment in two countries
S F La Vincente1, D Peel, S Carai
1Centre for International Child Health, Department of Paediatrics, University of Melbourne, Murdoch Childrens Research Institute, Royal Children's Hospital, Melbourne, Victoria, Australia. sophie.lavincente@rch.org.au
Setting:
The paediatric wards of hospitals in Malawi and Mongolia.
Objective:
To describe oxygen concentrator functioning in two countries with widespread, long-term use of concentrators as a primary source of oxygen for treating children.
Design:
A systematic assessment of concentrators in the paediatric wards of 15 hospitals in Malawi and nine hospitals in Mongolia.
Results:
Oxygen concentrators had been installed for a median of 48 months (interquartile range [IQR] 6-60) and 36 months (IQR 12-96), respectively, prior to the evaluation in Malawi and Mongolia. Concentrators were the primary source of oxygen. Three quarters of the concentrators assessed in Malawi (28/36) and half those assessed in Mongolia (13/25) were functional. Concentrators were found to remain functional with up to 30 000 h of use. However, several concentrators were functioning very poorly despite limited use. Concentrators from a number of different manufacturers were evaluated, and there was marked variation in performance between brands. Inadequate resources for maintenance were reported in both countries.
Conclusion:
Years after installation of oxygen concentrators, many machines were still functioning, indicating that widespread use can be sustained in resource-limited settings. However, concentrator performance varied substantially. Procurement of high-quality and appropriate equipment is critical, and resources should be made available for ongoing maintenance.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...