Gamma camera intrinsic uniformity in an unstable power supply environment
John E Ejeh1, Kayode S Adedapo, Bidemi I Akinlade
1Department of Nuclear Medicine, University College Hospital, University of Ibadan, P.M.B. 5116 Dugbe Post Office,Ibadan, South West Nigeria. jenyiejeh@yahoo.com
Abstract:
The main objective of this work was to show that a gamma camera in a developing country could perform efficiently despite electricity outages using intrinsic flood uniformity tests as an index of performance. A total of 143 intrinsic uniformity test results for a new gamma camera in use in an environment with unstable power supply are presented. The integral uniformity for the central field of view (CFOV) was found to be between 3.43% and 1.49% (3.29% for acceptance test) while the integral uniformity for the useful field of view (UFOV) was between 4.51% and 1.9% (5.21% for acceptance test). The differential uniformity for the CFOV was between 1.99% and 1.04% (2.25% for acceptance test) while that of the UFOV was between 2.84% and 1.23% (2.63% for acceptance test). In conclusion, these results show that the uniformity of the gamma camera under this condition is within an acceptable range for both planar and SPET imaging.