Establishment of dose reference levels for nuclear medicine in Greece
S Vogiatzi1, P Kipouros, M Chobis
1Greek Atomic Energy Commission, PO Box 60092, 15310 Agia Paraskevi, Greece. sbogiatz@eeae.gr
Abstract:
Greek Atomic Energy Commission's Department of Licensing and Inspections conducted a national survey for the establishment of nuclear medicine (NM) dose reference levels (DRLs) for adult patients, in Greece. The administered activities (AAs) (MBq) were collected from 120 NM departments (88 % of total), during on-site inspections for licensing purposes. Factors influencing the image quality were also investigated. The established national DRLs represent the AA value corresponding to the 75th percentile of the AA frequency distributions. In their majority, national DRLs and average AAs are comparable with the ones published in the international literature. In the light of new technologies, there might be potential for reducing the higher values of AAs, in co-operation with the nuclear medicine experts.