Performance of medical radiographic X-ray systems in Greece for the time period 1998-2004
S Economides1, C J Hourdakis, N Kalivas
1Licensing and Inspections, Greek Atomic Energy Commission, Ag. Paraskevi 153 10, Greece. sikonom@gaec.gr
Abstract:
This study presents the results of the on-site inspections performed by the Greek Atomic Energy Commission (GAEC) on conventional X-ray systems, both in public and private medical radiology departments. A part of the inspection concerns the assessment of important radiographic parameters obtained according to a specified quality control protocol and the comparison of the measured parameter values with the corresponding acceptance limits. A total number of 1011 radiographic systems were inspected by the GAEC during the period 1998-2004, with 63.4% of them being privately owned. Analysis of 8 different operational parameters is carried out providing information on the overall performance, as well as on each parameter of the inspected X-ray systems. Tube voltage reproducibility values show the highest percentage of acceptability (98.9%, 99.5% for private and public owned radiographic systems respectively), while linearity of radiation output for private systems (72.5%) and time accuracy for public ones (72.7%) show the worst results. The comparison of the results for the private sector to those of a similar study carried out during the period 1995-1997 indicates a substantial improvement in X-ray systems performance. Higher level of improvement shows exposure time accuracy (12.2% percentile increase) and linearity of radiation output (12.5% percentile increase). Nevertheless, the situation can be further optimized if maintenance and quality control of the radiographic systems are carried out on a more regular basis.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...