Reflectance measurement using digital camera and a protecting dome with built in light source
Related Experiment Video
Updated: May 30, 2026

Diffuse Reflectance Spectroscopy: Getting the Capillary Refill Test Under One's Thumb
Published on: December 2, 2017
Petri Välisuo1, Toni Harju, Jarmo Alander
1Faculty of Technology, Electrical and Energy Engineering, University of Vaasa, P.O. Box 700, 65101 Vaasa, Finland. petri.valisuo@uwasa.fi
Abstract:
The reflectance of the skin reveals the chemical and physical changes of the skin as well as many metabolic changes. The reflectance measurement is an important method for medical diagnosis, follow-up and screening. This article concentrates on designing and validating an imaging system, based on a digital camera. The proposed system can measure the reflectance of the skin with high spatial and currently four channel spectral resolution, in the range of 450 nm to 980 nm. The accuracy of the system is determined by imaging a colour checker board and comparing the obtained values with both given values and spectrometer measurements. The diffuse interreflections of both, the integrating sphere and the lighting dome of the imaging system, is compensated with a correction factor. The accuracy of the proposed system is only slightly weaker than the spectrometer. The imaging system characteristics are independent of the camera characteristics.