Transthoracic needle biopsy using a C-arm cone-beam CT system: diagnostic accuracy and safety
1Department of Radiology, School of Medicine, Ewha Womans University, Seoul, Republic of Korea.
Objective:
The purpose of this study was to evaluate the diagnostic accuracy and safety of performing transthoracic needle biopsy (TNB) under combined fluoroscopy and CT guidance using a C-arm cone-beam CT (CBCT) system.
Methods:
We evaluated the diagnostic accuracy and safety of performing TNB using a C-arm CBCT system. We retrospectively evaluated 99 TNB cases performed in 98 patients using a C-arm CBCT system with an 18-gauge automated cutting needle. We reviewed the diagnostic accuracy according to the size and depth of the lesion, incidence of complications, additional treatment for complications, procedure time, number of needle passes per biopsy and radiation dose.
Results:
The final diagnoses revealed 72 malignant and 27 benign lesions. The overall malignancy sensitivity, malignancy specificity and diagnostic accuracy were 95.8%, 100% and 97.0%, respectively, and those for small pulmonary nodules <20 mm in size were 94.1%, 100% and 96.6%, respectively. There was no significant difference in the correct diagnosis of malignancy according to lesion size (p = 0.634) or depth (p = 0.542). For benign lesions, a specific diagnosis was obtained in 14 cases (51.9%). TNB induced complications in 19 out of 99 procedures (19.2%), including pneumothorax (16.2%), immediate haemoptysis (2.0%) and subcutaneous emphysema (1.0%). Among these, four patients with pneumothorax required chest tube insertion (2.0%) or pig-tail catheter drainage (2.0%). The mean procedure time, number of needle passes and radiation doses were 11.9 ± 4.0 min, 1.2 ± 0.5 times and 170.0 ± 67.2 mGy, respectively.
Conclusion:
TNB using a C-arm CBCT system provides high diagnostic accuracy with a low complication rate and a short procedure time, particularly for small pulmonary nodules.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
04:10Robotic-assisted Bronchoscopy Combined with Multimodal Imaging for Targeted Lung Cryobiopsies
Published on: July 19, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...