Related Experiment Video
Updated: May 30, 2026

A Model to Simulate Clinically Relevant Hypoxia in Humans
Published on: December 22, 2016
Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in
Rhonda J Rosychuk1, Ann Hudson-Mason, Dianne Eklund
1Department of Paediatrics, Stollery's Children Hospital, University of Alberta, Edmonton, Alberta, Canada.
Background:
Discrepancies between pulse oximetry saturation (SpO(2)) and arterial saturation (SaO(2)) at low blood oxygenation values have been previously reported with significant variations among instruments and studies. Whether pulse oximeters that attenuate motion artifact are less prone to such discrepancies is not well known.
Objective:
To prospectively assess the agreement of the Masimo V4 pulse oximeter within the critical 85-95% SpO(2) target range.
Patients And Methods:
For all consecutive babies with gestational age <33 weeks, postnatal age <7 days, and an umbilical arterial line, SpO(2) was measured continuously and SaO(2) analyzed on an as-needed basis. Bland-Altman techniques provided estimates of the difference (D = SaO(2) - SpO(2)), standard deviation (SD), and 95% limits of agreement (D ± 2*SD).
Results:
There were 1,032 measurements (114 babies) with SpO(2) between 85 and 95%. The 95% limits of agreement were -2.0 ± 5.8, -2.4 ± 9.2, and -1.9 ± 5.3 in the SpO(2) categories 85-95, 85-89, and 91-95%, respectively. For the SpO(2) categories 85-89% and 91-95%, only 52% (53/101) and 59% (523/886) of SpO(2) values, respectively, corresponded to the analogous SaO(2) categories. In the 85-89% SpO(2) category, SaO(2) was lower than 85% in 39 of the 101 (39%) measurements.
Conclusion:
SaO(2) was lower on average than SpO(2) with an increased bias at lower saturation. The -2.4 ± 9.2 95% limits of agreement for SaO(2) - SpO(2) in the 85-89% SpO(2) category suggest that SpO(2) and SaO(2) are not interchangeable and intermittent SaO(2) assessments are warranted when the targeted SpO(2) is within this range.
More Related Videos
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...