Needle image plates compared to conventional CR in chest radiography: is dose reduction possible?
Vanessa Berger-Kulemann1, Sarah Pötter-Lang, Michael Gruber
1Department of Radiology, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. vanessa.berger-kulemann@meduniwien.ac.at
Purpose:
To compare image quality of standard-dose computed radiography and dose reduced needle-technology CR for supine CXR in a clinical setting.
Materials And Methods:
We prospectively evaluated 128 radiographs of 32 immunocompromised patients. For each patient four clinical CXR were performed within one week, two with powder image plates (PIP; Fuji ST-V) and two with needle image plates (NIP; Agfa DXS) at standard and half dose, respectively. One experienced radiologist and two residents blinded to dose level and kind of imaging system rated different anatomical structures, image noise, tubes/lines and abnormalities on a image quality scale from 1 to 10 (1=poor, 10=excellent). The rating scores were tested for statistical differences using analysis of variance with repeated measures.
Results:
A statistical difference (p<0.05) was found for the two systems as well as for the two dose levels. Overall rating scores were 6.5 for PIP with full dose, 6.2 for PIP with half dose, 7.6 for NIP with full dose and 7.4 for NIP with half dose. There was a significant difference in favour of the NIP system at the same dose level. Also the NIP images obtained at half dose were ranked significantly better compared to the PIP images at standard dose. The differences in ranking of anatomical structures and abnormalities were more pronounced in low absorption areas (pulmonary vessels, parenchyma) than in high absorption areas (mediastinum, spine).
Conclusion:
For supine chest radiograms the NIP technology allows for a dose reduction of 50% while providing higher image quality.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...