Surface dose measurement for helical tomotherapy
Related Experiment Video
Updated: May 30, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Jonatan A Snir1, Homeira Mosalaei, Kevin Jordan
1London Regional Cancer Program, London Health Sciences Centre, London, Ontario N6A 4L6, Canada.
Purpose:
To compare the surface dose measurements made by different dosimeters for the helical tomotherapy (HT) plan in the case of the target close to the surface.
Methods:
Surface dose measurements in different points for the HT plan to deliver 2 Gy to the planning target volume (PTV) at 5 mm below the surface of the cylindrical phantom were performed by radiochromic films, single use metal oxide semiconductor field-effect transistor (MOSFET) dosimeters, silicon IVD QED diode, and optically stimulated luminescence (OSL) dosimeters.
Results:
The measured doses by all dosimeters were within 12 +/- 8% difference of each other.
Conclusions:
Radiochromic films, EBT, and EBT2, provide high spatial resolution, although it is difficult to get accurate measurements of dose. Both the OSL and QED measured similar dose to that of the MOSFET detectors. The QED dosimeter is promising as a reusable on-line wireless dosimeter, while the OSL dosimeters are easier to use, require minimum setup time and are very precise.