Evaluation of objective test techniques in cochlear implant users with inner ear malformations
Betul Cicek Cinar1, Ahmet Atas, Gonca Sennaroglu
1Meders Medical, Ankara, Turkey. betulcicek01@gmail.com
Summary
Electrically evoked auditory brainstem response (EABR) is more effective than ECAP and ESRT for evaluating cochlear implant users with inner ear malformations. This study compared objective test techniques in these patients.
Area of Science:
- Otolaryngology
- Audiology
- Biomedical Engineering
Background:
- Inner ear malformations present unique challenges for cochlear implant (CI) fitting and evaluation.
- Objective test techniques are crucial for assessing auditory nerve function in CI users, especially those with complex inner ear anatomy.
Purpose of the Study:
- To compare the efficacy of objective electrophysiological tests, including electrically evoked compound action potentials (ECAPs), electrically evoked stapedius reflex thresholds (ESRTs), and electrically evoked auditory brainstem response (EABR) thresholds.
- To determine the most suitable objective test technique for evaluating and fitting cochlear implants in individuals with inner ear malformations.
Main Methods:
- A retrospective clinical study comparing CI users with inner ear malformations (n=20) to a control group with normal cochleae (n=15).
- Electrophysiological tests (ECAP, ESRT, EABR) were performed on 6 intracochlear electrodes after standard CI fitting.
- Statistical analysis included Mann-Whitney U test, Kruskal-Wallis analysis, and Dunn's Z test.
Main Results:
- Significantly lower percentages of acquired ECAP (25% vs. 74%) and ESRT (17.5% vs. 90%) thresholds were observed in the inner ear malformation group compared to the control group.
- Statistically significant differences were found in current levels and latencies of EABR wave V between the two groups.
- EABR demonstrated a higher applicability in the inner ear malformation group.
Conclusions:
- Electrically evoked auditory brainstem response (EABR) is a more applicable objective test technique for cochlear implant users with inner ear malformations than ECAP and ESRT.
- Findings suggest EABR should be prioritized for evaluation and fitting in this patient population.