Development of a performance verification test for USP apparatus 4
Joseph W Eaton1, Daren Tran, Walter W Hauck
1US Pharmacopeial Convention, 12601 Twinbrook Parkway, Rockville, Maryland, 20852-1790, USA. jwe@usp.org
Purpose:
To evaluate salicylic acid tablets as a candidate reference material in a Performance Verification Test (PVT) when a USP performance test for dissolution (General Chapter <711>) relies on USP Apparatus 4 (flow-through cell).
Methods:
We developed a dissolution procedure relying on Apparatus 4 and salicylic acid tablets. Thereafter, a designed experiment was conducted to identify operational variables that significantly affect salicylic acid dissolution in this apparatus.
Results:
Four variables (size of glass beads, cell temperature, flow rate, level of deaeration) and one combination effect (deaeration/bead size) were significant for mean percent dissolved. Two variables (tablet orientation, level of deaeration) were significant for standard deviation results, but these effects were less pronounced than those for mean percent dissolved results. Three variables (analyst, tester manufacturer, amount of glass beads) had no statistically significant effects on either the mean or standard deviation of the responses.
Conclusions:
The proposed PVT is capable of probing effects of changes in several critical operational parameters of Apparatus 4. Salicylic acid tablets were shown to be a suitable reference material for the PVT. The PVT using salicylic acid tablets satisfies important aspects of a PVT.
More Related Videos
07:15A Testing Platform for Durability Studies of Polymers and Fiber-reinforced Polymer Composites under Concurrent Hygrothermo-mechanical Stimuli
Published on: December 11, 2014
06:46A Performance-testing Platform for a Conduction Micropump with an FR-4 Copper-clad Electrode Plate
Published on: October 9, 2017
Related Concept Videos
In Vitro Drug Dissolution: Compendial Testing Models I
In Vitro Drug Dissolution: Compendial Testing Models II
In Vitro Drug Release Testing: Overview, Development and Validation
Testing Water Quality
Data Validation
Key parameters for method validation include: