Could it be advantageous to tune the temperature controller during radiofrequency ablation? A feasibility study using
José Alba-Martínez1, Macarena Trujillo, Ramón Blasco-Giménez
1Biomedical Synergy, Electronic Engineering Department, Universtat Politècnica de València, Spain.
Tailoring proportional-integral (PI) controller parameters for radiofrequency (RF) cardiac ablation did not significantly alter lesion size or maximum temperature. A standard PI controller may perform adequately across various conditions.
Area of Science:
- Biomedical Engineering
- Medical Devices
- Computational Modeling
Background:
- Radiofrequency (RF) cardiac ablation is a key procedure for treating arrhythmias.
- Controller performance in RF ablation is critical for lesion efficacy and safety.
- Optimizing controller parameters (Kp and Ki) may enhance dynamic behavior.
Purpose of the Study:
- To evaluate if tuning proportional-integral (PI) controller parameters (Kp and Ki) improves dynamic control during RF cardiac ablation.
- To determine if customized PI controllers can accelerate control action and increase lesion size.
Main Methods:
- Developed theoretical models solved using the finite element method.
- Simulated RF cardiac ablations with temperature controlled at 55°C.
- Implemented PI controllers with parameters adapted to varying tissue properties, electrode-tissue contact, and electrode characteristics.
Main Results:
- Lesion dimensions and maximum temperature (Tmax) showed minimal changes (<1.9°C, <0.2 mm width, <0.3 mm depth) with a tailored PI controller compared to a standard one.
- A direct relationship was observed between controller response time and transient electrode temperature.
Conclusions:
- A PI controller optimized for standard conditions may provide benefits even with variations in tissue, electrode-tissue contact, and electrode characteristics.
- The study suggests robustness of standard PI controller tuning in RF cardiac ablation.