Retention cuff pressure study of 3 indwelling stool management systems: randomized study of 10 healthy subjects
Floriano Marchetti1, Joseph P Corallo, Jaime Ritter
1University of Miami Miller School of Medicine, University of Miami Hospital, Miami, Florida, USA.
Purpose:
The purpose of this study was to compare retention cuff pressures of 3 indwelling stool management systems while subjects assumed different body positions and while cuffs were inflated to different volumes.
Methods:
Retention cuff pressure study of 3 indwelling stool management systems was a randomized, crossover, open-label pilot study of 10 healthy adult volunteers in 3 body positions (supine, right side, and left side), 3 head-of-bed elevations (20°, 30°, and 40°), and 3 cuff overfill volumes (5, 10, and 15 mL). The devices were DigniCare Stool Management System (device A; Bard Medical Division, C. R. Bard, Inc, Covington, Georgia), Flexi-Seal Fecal Management System (device B; ConvaTec, a division of E. R. Squibb & Sons, LLC, Princeton, New Jersey), and ActiFlo Indwelling Bowel Catheter System (device C; Hollister, Inc, Libertyville, Illinois). We assessed cuff pressure by manometry and rectal mucosa by digital examination and small-diameter, flexible endoscopy.
Results:
Cuffs were appropriately seated in the rectal vault for all 3 devices in all body positions and overfill volumes. Rectal mucosal abnormalities were observed in 4 of 10 subjects (40%) after removal of device A, 1 of 5 (20%) after removal of device B, and 3 of 5 (60%) after removal of device C. Retention cuff pressure was at least 2-fold lower for device A than for device B or C in all body positions, head-of-bed elevations, and device overfill volumes. For example, mean pressure while subjects were on their left sides was 25.0 mm Hg for device A, 79.2 mm Hg for device B, and 67.2 mm Hg for device C. Corresponding pressures at 15 mL of overfill were 52.5, 102.0, and 94.0 mm Hg. Subject comfort scores were comparable for all 3 devices.
Conclusion:
All devices appeared to seat well within the rectal vault, but device A was associated with cuff pressure measurements that were consistently lower than those of devices B and C. More studies are needed to elucidate the clinical relevance of these findings and whether they translate to differences in patient safety or comfort.
Related Concept Videos
Inflammatory Bowel Disease V: Surgical Management
Here are some common surgical interventions for IBD:
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Drugs for Treatment of Constipation-Predominant IBS
Inflammatory Bowel Disease III: Diagnostic Studies and Management I-Nutritional Therapy
Diagnostic studies
A colonoscopy is the definitive screening test, distinguishing ulcerative colitis from other colon diseases with similar symptoms. During a colonoscopy test, inflamed mucosa with exudate ulcerations can be observed, and biopsies are taken to determine the histologic characteristics of the colonic...
Irritable Bowel Syndrome III: Medical and Nursing Management
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...