Related Experiment Video
Updated: May 29, 2026

An Automated System for Sound Localization Testing in Hearing-Impaired Listeners
Published on: March 13, 2026
Study on the applicability of instrumental measures for black-box evaluation of static feedback control in hearing
1ExpORL-Department of Neurosciences, K.U. Leuven, B-3000 Leuven, Belgium. nilesh.madhu@rub.de
Abstract:
Presented is a report on black-box evaluation of feedback control systems for commercial hearing aids. The aim of the study is to examine the ability of existing instrumental measures to quantify the performance of the feedback control system in black-box settings and on realistic signals, when more than one element of the signal processing chain may be active (compression, noise suppression, microphone directionality, etc.). The evaluation is carried out on 6 different hearing aids and for 10 measures. Thereby it is possible to see which measure is best suited to measuring which specific characteristic of the feedback control system, and serves as a beginning for conducting perceptual tests. The study uses static (but variable) feedback paths and is based on signals recorded from the in-ear microphone of an artificial head, on which the hearing instruments are mounted.
Related Concept Videos
Effects of feedback
Feedback significantly modifies the gain of a control system. The gain of a system without feedback is altered by a factor of one plus GH, where G represents...
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...