Quantitative determination of methylnaltrexone in human serum using liquid chromatography-tandem mass spectrometry
Stefan Oswald1, Gitta Schumacher, Werner Siegmund
1Department of Clinical Pharmacology, Ernst-Moritz-Arndt University Greifswald, Friedrich-Loeffler-Str. 23d, 17487 Greifswald, Germany. stefan.oswald@uni-greifswald.de
Abstract:
Methylnaltrexone (MNTX) is a novel peripherally acting μ-opioid antagonist that prevents peripheral side effects of opioid drugs such as constipation without affecting the analgesia. We developed a selective and sensitive assay to measure MTNX concentrations in human serum. The drug was measured after protein precipitation with perchloric acid using naltrexone as internal standard and liquid chromatography-tandem mass spectrometry (LC-MS/MS) for detection. The chromatography was performed isocratically on a RP18 column using 25 mM ammonium acetate buffer (pH 4)/acetonitrile (90%/10%; flow rate 200 μl/min) as mobile phase. The MS/MS analysis was performed in positive ionization mode monitoring the m/z transitions 356.4/284.2 for MNTX and 342.4/324.2 for naltrexone. The method was validated according to selectivity, linearity, accuracy, precision, recovery, matrix effects and stability. The validation range for MNTX in serum was 0.5-250 ng/ml. The developed LC-MS/MS was shown to be valid and successfully applied to measure serum-concentration-time curves of MNTX in a pilot study in healthy volunteers.