Related Experiment Video
Updated: May 29, 2026

Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
Correction of susceptibility-induced GRE phase shift for accurate PRFS thermometry proximal to cryoablation iceball
Antje Kickhefel1, Clifford Weiss, Joerg Roland
1Eberhard-Karls-University Tübingen, Tübingen, Badem-Württenberg, Germany. antje.kickhefel@siemens.com
Introduction:
The susceptibility contrast between frozen and unfrozen tissue disturbs the local magnetic field in the proximity of the ice-ball during cryotherapy. This effect should be corrected for in real time to allow PRFS-based monitoring of near-zero temperatures during intervention.
Material And Methods:
Susceptibility artifacts were corrected post-processing, using a rapid numerical algorithm. The difference in bulk magnetic susceptibility between frozen and non-frozen tissue was approximated to be uniform over the ice-ball volume and was determined from the isothermal principle applied to the phase-transition frontier of compartments. Subsequently, the magnetic perturbation field was calculated rapidly in 3D using a Fourier-convolution. Experimental studies were performed for two scenarios: tissue defrosting in a water bath and induction of an ice-ball by a MR-compatible cryogenic probe.
Results:
The susceptibility artifacts yielded PRFS temperature errors as high as 10-12°C proximal to the ice-ball, positive or negative depending on the relative orientation of the position vector from the B(o) direction. These effects were fully corrected for to within the noise range. The susceptibility-corrected PRFS temperature values were consistent with the phase-transition isothermal condition, irrespective of the local orientation of the position vector.
Conclusion:
By implementing on-line the post processing algorithm, PRFS MRT may be used as a safety tool for non-invasive and accurate monitoring of near-zero temperatures during MR-guided clinical cryotherapy.