A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Activity quantification combining conjugate-view planar scintigraphies and SPECT/CT data for patient-specific 3-D
Yannick Berker1, Andreas Goedicke, Gerrit J Kemerink
1Department of Experimental Molecular Imaging, RWTH Aachen University Hospital, Pauwelsstraße 30, 52074 Aachen, Germany. yberker@ukaachen.de
Purpose:
Three-dimensional dosimetry based on quantitative SPECT/CT has potential advantages over planar approaches, but may be impractical due to acquisition durations. We combine one SPECT/CT with improved quantification of multiple planar scintigraphies to shorten acquisitions.
Methods:
A hybrid 2-D/3-D quantification technique is proposed, using SPECT/CT information for robust planar image quantification and creating virtual SPECTs out of conjugate-view planar scintigraphies; these are included in a 3-D absorbed dose calculation. A projection model simulates photon attenuation and scatter as well as camera and collimator effects. Planar and SPECT calibration techniques are described, offering multiple pathways of deriving calibration factors for hybrid quantification. Model, phantom and patient data are used to validate the approach on a per-organ basis, and the similarity of real and virtual SPECTs, and of planar images and virtual SPECT projections, is assessed using linear regression analysis.
Results:
Organ overlap, background activity and organ geometry are accounted for in the algorithm. Hybrid time-activity curves yield the same information as those derived from a conventional SPECT evaluation. Where correct values are known, hybrid quantification errors are less than 16% for all but two compartments (SPECT/CT 23%). Under partial volume effects, hybrid quantification can provide more robust results than SPECT/CT. The mean correlation coefficient of 3-D data is 0.962 (2-D 0.934). As a consequence of good activity quantification performance, good agreement of absorbed dose estimates and dose-volume histograms with reference results is achieved.
Conclusion:
The proposed activity quantification method for 2-D scintigraphies can speed up SPECT/CT-based 3-D dosimetry without losing accuracy.