Determining of Ablation Zone in Ex Vivo Bovine Liver Using Time-Shift Measurements
Mohamed Lamhamdi1, Ali Esmaeili1, Kiyan Layes1
1Deparment of Engineering, RheinMain University of Applied Science, 65428 Rüsselsheim, Germany.
Abstract:
This study presents a measurement principle for determining the size of the ablation zone in MWA, which could ultimately form an alternative to more expensive monitoring approaches like CT. The measurement method is based on a microwave transmission measurement. A MWA is performed experimentally on ex vivo bovine liver to determine the ablation zone. This setup uses a custom slot applicator performing the MWA at an operating frequency of 2.45 GHz and a custom bowtie antenna measuring the waves transmitted from the applicator. Furthermore, a custom measurement probe is used to determine the dielectric properties. A time-shift analysis is used to determine the radial extent of the ablation zone. Several measurements are carried out with a power of 50 W for 10 min to show the reproducibility. The results show that this method can provide reproducible outcomes to determine the ablation zone with a maximum error of 4.11%.