Continuous-flow analysis of ammonia in perchloric acid supernate of blood or plasma using an ammonia-selective
Abstract:
A simple automated method for the estimation of ammonia in perchloric acid supernate of blood or plasma using an ion-selective electrode (Orion Ammonia-selective electrode, Model 95-10) is described. The reliability of the proposed method has been checked against an ion-exchange resin procedure, which has been chosen as a standard procedure. Regression equation and correlation coefficient for the proposed method are y = 0.7 X + 10 and 0.945, respectively, as compared with the chosen standard method. Within-run and between-run precision are 2.1% and 3.5% respectively. The average percent recovery is 97.5% and a tentative range is 13-73 microgram/dl (9-52 micronmol/l) ammonia nitrogen.
Related Concept Videos
Titration of a Weak Base with a Strong Acid
Using the ICE table and substituting the Kb value, we calculate the initial pH of 50 mL of 0.1 M ammonia to be 11.11. Addition of 25 mL of 0.1 M hydrochloric acid to this solution of ammonia results in a buffer with an equal concentration of ammonia and ammonium ions. The pH of this buffer can be calculated by substituting these...