Background And Aims:
Every country has standardized reprocessing guidelines for reducing the risk of microorganism transmission via reusable biopsy forceps. Sterilization is performed either by autoclaving or with the use of ethylene oxide (EO) gas. However, there are no clear standard global recommendations. The aim of this study was to determine whether EO gas or autoclaving is a safer and more effective method for the sterilization of reusable forceps.
Methods:
This was a prospective study conducted at multiple tertiary referral centers. Seventy reusable biopsy forceps that had been reused at least 20 times each were collected from six endoscopy centers. In all, 61 forceps from five centers were sterilized using EO gas, and the nine forceps from the remaining center were placed in an autoclave. We performed real-time polymerase chain reaction (RT-PCR) for Mycobacterium tuberculosis and hepatitis B virus and performed bacterial cultures on the reusable forceps, which were cut into 2- to 3-cm sections. The forceps were also scanned with an electron microscope (EM) to detect surface damage and contamination.
Results:
Escherichia coli bacteria were cultured from 2 of the 61 (3.3%) reusable biopsy forceps sterilized with EO gas. On EM scanning, abundant debris and tissue materials remained on the cup surfaces of the reused biopsy forceps and on their inner wires. No microorganisms were found on the autoclaved forceps.
Conclusions:
Sterilization with EO gas may be inadequate because the complicated structure of the forceps may interfere with sterilization. Therefore, for optimum safety, reusable biopsy forceps should be sterilized by autoclaving.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...