Related Experiment Video
Updated: May 29, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
[Simultaneous determination of trihexyphenidyl, chlorpromazine and clozapine in blood by GC-MS]
Shui-Qing Zheng1, Wei Wang, Chen Liang
1Shanghai Key Laboratory of Crime Scene Evidence, Institute of Forensic Science, Shanghai Public Security Bureau, Shanghai 200083, China.
Objective:
To develop a method to measure trihexyphenidyl, chlorpromazine and clozapine in human blood with GC-MS.
Methods:
The specimens were alkalized (pH > 10) and extracted with V (benzene):V(ethyl acetate) = 1:1, and qualitatively analyzed using GC-MS-Full Scan with internal standard SKF525A. The specimens were alkalized (pH > 10) and extracted with V(benzene):V(ethyl acetate) = 1:1, and quantitatively analyzed using GC-MS-SIM with internal standard diazepam-d5.
Results:
The lowest detection limits of trihexyphenidyl, chlorpromazine and clozapine were 0.3, 0.3 and 0.7 ng/mL (S/N > or = 3) respectively. The calibration curve in 20-10 000 ng/mL showed a good linear distribution. The recovery rate was 79.9% to 85.5%. The RSDs of intraday and interday were less than 5.1%.
Conclusion:
The established method was simple, sensitive and accurate for simultaneous determination of trihexyphenidyl, chlorpromazine and clozapine in human blood, and can be applied in forensic toxicological cases.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...