Multicentre evaluation of the Boehringer Mannheim/Hitachi 717 analysis system.
H Baadenhuijsen1, P M Bayer, H Keller
1Radboud Ziekenhuis, Nijmegen, The Netherlands.
Summary
The Boehringer Mannheim/Hitachi 717 analysis system demonstrates excellent photometric performance and precision for clinical chemistry. This multicentre trial confirmed its reliability and ease of use in laboratory settings.
Area of Science:
- Clinical Chemistry
- Laboratory Instrumentation
- Analytical Performance Evaluation
Background:
- The introduction of new automated clinical chemistry analyzers is crucial for improving laboratory efficiency.
- Evaluating the performance of novel diagnostic equipment is essential before widespread adoption.
Purpose of the Study:
- To assess the analytical performance of the Boehringer Mannheim/Hitachi 717 analysis system.
- To validate the system's precision, linearity, and carryover in a multicentre setting.
Main Methods:
- European multicentre trial involving the Hitachi 717 system.
- Photometric linearity assessment up to an absorbance of 2.8.
- Photometric imprecision evaluation using specific wavelength pairs.
- Within-run and between-day imprecision analysis for 13 analytes.
- Correlation studies with existing analytical instruments.
- Assessment of instrument drift and carryover.
Main Results:
- Photometer linearity confirmed up to 2.8 absorbance.
- Maximal photometric imprecision (CV) was 0.5% at 340/405 nm (absorbance 0.9-2.4).
- Mean within-run imprecision < 2%; mean between-day imprecision < 2.5% for 13 analytes.
- High correlation with comparison instruments and linearity exceeding manufacturer claims.
- No significant drift or carryover detected under experimental conditions.
Conclusions:
- The Hitachi 717 system exhibits high analytical performance and reliability.
- The instrument is user-friendly with simple daily maintenance, facilitating easy adoption.
- This system is a valuable addition for clinical chemistry laboratories.