Physical Activity Measurement in Children Accepting Table Tennis Training
Published on: July 27, 2022
Validity and Comparability of a Wrist-Worn Accelerometer in Children
Orjan Ekblom1, Gisela Nyberg, Elin Ekblom Bak
1Dept of Clinical Sciences, Intervention, and Technology, Stockholm, Sweden.
Journal of Physical Activity & Health
|September 22, 2011
Summary
The wrist-worn Actiwatch is a valid and reliable tool for measuring children's energy expenditure and physical activity intensity. This study established cut-off values for activity intensity, aiding accurate assessment in youth.
Area of Science:
- Pediatric Exercise Science
- Biomedical Engineering
- Physical Activity Measurement
Background:
- Wrist-worn accelerometers offer a practical alternative to hip-worn devices for long-term physical activity monitoring.
- Assessing physical activity in children is crucial for understanding health behaviors and outcomes.
Purpose of the Study:
- To validate the Actiwatch (wrist-worn) for estimating energy expenditure in children.
- To establish cut-off values for classifying light, moderate, and vigorous physical activity.
- To compare the Actiwatch with the Actigraph (hip-worn) and assess the Actiwatch's reliability.
Main Methods:
- Indirect calorimetry served as the criterion measure for validity assessment.
- Receiver Operating Characteristic (ROC) analyses were used to determine activity intensity cut-off values.
- Simultaneous wear of Actiwatch and Actigraph during free-living conditions assessed comparability; reliability was tested using a mechanical shaker.
Main Results:
- A significant correlation (0.80) was found between accelerometer output and energy expenditure.
- Cut-off values for 1.5, 3, and 6 METs were identified as 80, 262, and 406 counts per 15 seconds, respectively.
- Energy expenditure estimates showed significant differences between Actiwatch and Actigraph; Actiwatch demonstrated low intra- and inter-instrument variation (0.72%–8.4%).
Conclusions:
- The wrist-worn Actiwatch is a valid and reliable instrument for estimating energy expenditure and physical activity intensity in children aged 8–10 years.
- Established cut-off values facilitate accurate classification of physical activity intensity using the Actiwatch in pediatric populations.