Related Experiment Video
Updated: May 29, 2026

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Estimating core temperature with external devices after exertional heat stress in thermal protective clothing
Riana R Pryor1, Jennifer R Seitz, Julia Morley
1Department of Emergency Medicine, Emergency Responder Human Performance Laboratory, University of Pittsburgh, 3600 Forbes Avenue, Suite 400A, Pittsburgh, PA 15261, USA.
Background:
Temperature measurement is important for emergency medical services (EMS) providers when identifying and treating heat illness or infection. Direct measures of body core temperature (T(c)) are often expensive (ingestible capsules) or impractical (rectal probes) in the field. Multiple devices for estimating T(c) have been adopted by EMS providers, with little understanding of the agreement between these devices and T(c).
Objective:
To examine the agreement between the results of five external thermometers and T(c) after subjects experienced physical exertion while wearing protective clothing.
Methods:
Fifty firefighters completed treadmill walking in thermal protective clothing in a hot environment. Measurements of core, temporal, tympanic, forehead, and skin temperatures were obtained during a 20-minute recovery period simulating emergency incident rehabilitation.
Results:
The mean bias of external thermometers ranged from -1.31°C to -3.28°C when compared with T(c) and exceeded the predetermined clinical cutoff of ±0.5°C from T(c). The 95% limits of agreement ranged from 2.75°C to 5.00°C.
Conclusions:
External measuring devices failed to accurately predict T(c) in hyperthermic individuals following exertion. Confidence intervals around the bias were too large to allow for reasonable estimation of T(c). EMS providers should exercise caution when using any of these temperature estimation techniques.
Related Concept Videos
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Body Temperature