Use of a remotely steerable "robotic" catheter in a branched endovascular aortic graft
Tom Carrell1, Neville Dastur, Richard Salter
1King's Health Partners, London, United Kingdom; Guy's and St Thomas' Hospitals NHS Foundation Trust, London, United Kingdom; King's College London, London, United Kingdom. tom.carrell@gstt.nhs.uk
Abstract:
We report the use of a remotely steerable catheter to treat kinked renal bridging stents 8 months after branched endovascular repair of a type III thoracoabdominal aortic aneurysm. Conventional techniques using single, coaxial, and manually steerable sheaths proved too unstable to provide the support required to pass a wire against resistance through the kinked stent. A remotely steerable "robotic" catheter provided sufficient precision and stability to cross the kink and reline it with an additional stent, restoring renal perfusion. This technology can help achieve precise and stable introducer sheath position. Further evaluation is necessary to understand the wider applications.