Related Experiment Video
Updated: May 28, 2026

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesis
J U Lindgren1, B H Brismar, A C Wikstrom
1Karolinska University Hospital, Department of Orthopaedics K54, Huddinge, S14186 Stockholm, Sweden. urban.lindgren@ki.se
Abstract:
A 70-year-old man with an uncemented metal-on-polyethylene total hip prosthesis underwent revision arthroplasty 33 months later because of pain, swelling and recurrent dislocation. There appeared to be corrosion and metal release from the prosthetic head, resulting in pseudotumour formation and severe local soft-tissue destruction. The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.