Small mass-change detectable quartz crystal microbalance and its application to enzymatic one-base elongation on DNA
Hiroshi Yoshimine1, Taisuke Kojima, Hiroyuki Furusawa
1Department of Biomolecular Engineering, Tokyo Institute of Technology and SENTAN, Midori-ku, Yokohama, Japan.
Abstract:
A highly sensitive 27 MHz quartz crystal microbalance instrument with an automatic flow injection system was developed to obtain realistic minimal frequency noise (±0.05 Hz) and to obtain a stable signal baseline (±1 Hz/h) by controlling the temperature of each part in the quartz crystal microbalance (QCM) system using three Peltier devices with a resolution of ±0.001 °C and by optimizing the flow system to prevent fluctuation of the internal pressure of the QCM. The improved QCM with an automatic flow injection system enabled detection of small mass changes such as binding of biotin to a streptavidin-immobilized QCM with a high signal-to-noise ratio. We also applied this device to enzyme reactions of one-base elongation by DNA polymerase (Klenow fragment, KF). We immobilized dsDNAs including the protruding end of dA, dG, dT, or dC on the QCM electrode and ran complementary dNTP monomers with KF into the QCM flow cell. We could directly detect the enzymatic one-base elongation of DNA as a small mass increase, and we found the difference in the reaction rate for each monomer.
More Related Videos
06:24Real-time Observation of the DNA Strand Exchange Reaction Mediated by Rad51
Published on: February 13, 2019
07:11Dissipative Microgravimetry to Study the Binding Dynamics of the Phospholipid Binding Protein Annexin A2 to Solid-supported Lipid Bilayers Using a Quartz Resonator
Published on: November 1, 2018