Limitations of anaesthesia depth monitoring
Francisco A Lobo1, Stefan Schraag
1Anaesthesiology Department, Hospital Geral de Santo António - Centro Hospitalar do Porto, Porto, Portugal. xlanesth@gmail.com
Purpose Of Review:
We critically review brain function monitors based on the processed electroencephalogram with regards to signal quality, artefacts and other limitations in clinical performance.
Recent Findings:
Several studies have been showing that depth of anaesthesia monitors based on processed electroencephalogram has limitations that can lead to a wrong interpretation of the level of anaesthesia. Processed electroencephalogram indices can be altered by nonanaesthetic influences ranging from artefacts that affect signal quality and signal processing, adverse effects of some anaesthetic and nonanaesthetic drugs, neuromuscular blocking agents to conditions inherent to the patient such as cerebral tumours, brain ischemia and temperature.
Summary:
Clinicians should be aware of the several limitations of the commercial devices intending to monitor the depth of anaesthesia, which may not reflect the real underlying level of unconsciousness.
Related Concept Videos
Stages of General Anesthesia
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Parenteral Anesthetics: Overview
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Local Anesthetics: Clinical Application as Surface, Infiltration, and Conduction Block Anesthesia
Inhalational Anesthetics: Overview