Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Effect of insulation material in aging pacing leads: comparison of impedance and other electricals: time-dependent
W Ben Johnson1, Alan Braly, Kenneth Cobian
1Iowa Heart Center, PC, Des Moines, Iowa 50134, USA. wbjohnson@iowaheart.com
Background:
There has been concern over declining bipolar (BP) impedance (Z) in aging polyurethane (PU) cardiac pacing leads. Subsequently, a prospective study was conducted comparing BP Z, threshold (Th), and R-wave sensing amplitude of 55D PU-insulated (Model 4024, Medtronic, Inc., Minneapolis, MN, USA) and silicone-insulated (Model 5024) leads.
Methods:
This study was initiated by The Iowa Heart Center. Patients with Model 4024 (N = 162) or 5024 (N = 120) pacing leads with at least 6 years implant time were enrolled and followed for an additional 5 years.
Results:
There was a significant drop in the mean BP Z for the Model 4024 population, between enrollment (6 years) and the final endpoint (11 years), which was in contrast to the Model 5024 which did not see a significant drop in its mean BP Z for this same period. The trend difference seen in the means between the two models was statistically significant (P < 0.0001). In addition, a statistically significant relationship was found between dropping BP Z and rising Th (P < 0.0001). The analysis showed that if BP Z dropped below 200 ohms, the probability of having a >3X increase over baseline, in Th at 2.5 V, increases from approximately 3-7% to as high as 30%.
Conclusions:
A significant drop in BP Z observed in the PU-insulated Model 4024 lead was not present in the silicone-insulated Model 5024 lead. The statistically significant relationship between dropping BP Z and rising Th helps to understand how to better manage patients with aging leads.