Background:
Subxyphoid active left ventricular epicardial (LVE) lead implants or VT ablation are attractive but remain a challenge due to concerns of coronary artery damage. We aimed to see if Doppler-guided positioning could permit safe LVE lead placement without coronary angiography. We evaluated the feasibility of a Doppler flow-guided subxyphoid epicardial screw-in lead fixation in a swine model.
Methods:
Acute subxyphoid access to the pericardial space was performed in an anesthetized swine model using a deflectable sheath and a modified needle-derived Doppler flow meter. The audio signal and visual display from the Doppler flow meter were recorded. Coronary angiography was performed to verify the catheter location. A SelectSecure Model 3830 lead (Medtronic) was used to assess pacing in the procedure.
Results:
In both of two swine, the deflectable catheter was inserted into pericardial space via subxyphoid access. The tip of the deflectable catheter with the Doppler was directed to several locations, from quiet (no nearby coronary artery expected) to typical rhythmic pulsatile sound locations which were maximal when superimposed on a coronary artery. Repeated coronary angiograms confirmed the expected findings. A 3830 active lead was fixed into a quiet location for LVE pacing, and confirmed by angiography as distant from a coronary artery.
Conclusions:
Doppler-guided subxyphoid epicardial screw-in lead placement is feasible once the catheter tip is directed and stabilized in a desired LVE location. This obviates the need for repeated (or any) coronary angiography. The Doppler-guided subxyphoid epicardial procedure may also be applicable for epicardial ventricular arrhythmia ablation procedures.
More Related Videos
08:31A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization