Balancing image quality and compression factor for special stains whole slide images
Anurag Sharma1, Pinky Bautista, Yukako Yagi
1NEC Laboratories America, Inc., Princeton, NJ, USA. asharma@neclabs.com
Unlabelled:
The objective is to find a practical balance between quality and performance for daily high volume whole slide imaging. We evaluated whole slide images created by various scanners at different compression factors to determine the best suitable quality factor (QF) needed for pathological images of special stains.
Method:
We scanned two sets of eight special stains slides each at 0.50 μm/pixel resolution in Hamamatsu scanner at six and five QF levels respectively to generate 72 images which were observed at a calibrated monitor by imaging specialists, a histo-technician, and a pathologist to find the most suitable QF level for special stains in digital slides.
Results:
Most special stains images were acceptable at QF 30 except for the stain Reticulin where the lowest acceptable QF was 50. The compression of images from QF 90 to QF 50 reduced the size of the images by 62.73%.
Conclusion:
0.50 μm/pixel images at QF 50 or above were found suitable 12 special stain.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...