Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
Effect of alteration in pump speed on pump output and left ventricular filling with continuous-flow left ventricular
Christopher S Hayward1, Robert Salamonsen, Anne M Keogh
1Heart Failure and Transplant Unit, St. Vincent's Hospital, Sydney, New South Wales, Australia. cshayward@stvincents.com.au
Abstract:
Third-generation continuous-flow left ventricular assist devices (LVAD) provide reduced pulsatility flow. We examined the safe working range for LVAD pump speed and the effect on pump output and cardiac function in 13 stable outpatients with VentrAssist-LVAD (Ventracor Ltd, Australia). Pump speed was decreased from a baseline mean of 2,073 ± 86 revolutions per minute (RPM, with corresponding mean flow of 5.59 ± 1.18 L/min, mean ± standard deviation) to an average low-speed of 1,835 ± 55 RPM (corresponding flow 4.68 ± 0.99 L/min) and up to high-speed of 2,315 ± 66 RPM (corresponding flow 6.30 ± 1.29 L/min). There was a strong linear relationship between alteration in speed and flow rates (r(2) = 0.89, p < 0.00001) but marked interpatient variation. Downward titration to preset minimum 1,800 RPM was achieved in 9/13 (69%) and upward titration to the preset maximum 2,400 RPM was achieved in 4/13 (31%). Upward titration was stopped due to ventricular suction or nonsustained ventricular tachycardia (VT) in 4/13 (31%). Ventricular suction or VT (in 4/13) tended to be more common in patients with poor right ventricular (RV) function (p = 0.07). In summary, pump flow is stable within a relatively small speed range and should not be altered without close monitoring due to variation in response between patients, particularly with concomitant RV impairment.
More Related Videos
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Mitral Stenosis I: Introduction